Pressure Controller/Calibrator
Calibration Tables
103
Table 34. PM600-G100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 0
1
60 30
2 25 60 30
3 50 60 30
4 75 60 30
5 100 60 30
Descending Points
6 75 60 30
7 50 60 30
8 25 60 30
9 0 60 30
1. UUT is zeroed before setting pressure.
Table 35. PM600-A100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 6 60 30
2 25 60 30
3 50 60 30
4 75 60 30
5 100 60 30
Descending
Points
6 75 60 30
7 50 60 30
8 25 60 30
9 6 60 30