Pressure Controller/Calibrator
Maintenance
19
PMM and PCM Removal
To remove the PMMs, see Figure 7:
Caution
To prevent damage to the sensors inside the modules, do not
drop the modules.
1. Open the module bay as described in Access the Module Bay.
2. Turn the knob on the PMM counter-clockwise until it loosens.
3. Pull the PMM towards you and slide the module out.
To remove the PCM:
1. Follow step 1 from the previous procedure.
2. Use a 6 mm nut driver to loosen the two nuts on the PCM .
3. Pull the PCM towards you and slide the module out.