The Product has a fuse that protects from over current. See Table 3. Access the
fuse from the rear panel. The fuse-rating label below the fuse holder shows the
correct replacement fuse ratings for each operating voltage.
Warning
To prevent possible electrical shock, fire, or personal injury:
Turn the Product off and remove the mains power cord.
Stop for 2 minutes to let the power assemblies discharge
before you open the fuse door.
Use only specified replacement fuses, see Table 3.
To replace the fuse, see Figure 8:
1. Make sure the Product is off and disconnect the mains power cord.
2. With a standard screwdriver, release the fuse holder door.
3. Pull out the fuse holder.
4. If necessary, replace the fuse.
5. Reinsert the fuse holder and snap it shut.
Table 3. Replacement Fuses
Fuse Description Fluke Part Number
Fuse 5x20MM, 2 AMP 1297149