Figure 40. Calibrating Sensors
Supply Port (ATM and 3,000 psi)
Note
There is an isolation valve between the supply port on the manifold
and the supply port of the PCM. Follow directions to ensure that the
supply pressure and valve position is correct for the calibration point.
Send the CAL:PRES6:VAL1 xx.xx command to open the supply
isolation valve. This command will be used twice.
1. Remove all supply pressure from the supply port of the Product. See
Figure 41.
2. Use a TEE to connect an external digital pressure device that reads in
Absolute units to the supply line. Use this as the reference pressure.
3. Touch Vent and then Enter to vent the Product.
4. Ensure the supply line is disconnected from the supply port.
5. With the reference and supply sensor vented to atmosphere, type
CAL:PRES6:VAL1 xx.xx (xx.xx =value in units must be absolute). This