EasyManua.ls Logo

Fluke 6270A - PM600

Fluke 6270A
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pressure Controller/Calibrator
Calibration Adjustment
77
PM600
PM600 Range
Min
Pressure
Max
Pressure
Reference
System
1
Piston Cylinder
1
Reference Mode
Dwell
Time
Averaging
Time
------------ [kPa] [kPa] ------------- ------------- ------------- [s] [s]
A20M ATM 20 000 PG7202
200 kPa, 500 kPa, 1 MPa, 2
MPa /kg P-C
Absolute by ATM 60 30
A14M ATM 14 000 PG7202
200 kPa, 500 kPa, 1 MPa, 2
MPa /kg P-C
Absolute by ATM 60 30
A10M ATM 10 000 PG7102 100 kPa/kg P-C Absolute by ATM 60 30
A7M ATM 7000 PG7102 100 kPa/kg P-C Absolute by ATM 60 30
A3.5M 70 3500 PG7601 100 kPa/kg P-C Absolute by Vacuum 60 30
A2M 70 2000 PG7601 100 kPa/kg P-C Absolute by Vacuum 60 30
A1.4M 35 1400 PG7601 50 kPa/kg P-C Absolute by Vacuum 60 30
A700K 20 700 PG7601 20 kPa/kg P-C Absolute by Vacuum 60 30
A350K 10 350 PG7601 10 kPa/kg P-C Absolute by Vacuum 60 30
A200K 10 200 PG7601 10 kPa/kg P-C Absolute by Vacuum 60 30
G200K 0 200 PG7601 10 kPa/kg P-C Gauge 60 30
A100K 6 100
PG7607,
2465
5 kPa/kg P-C, (TL) P-C Absolute by Vacuum 60 30
BA100K 70 110
PG7607,
PG7601,
2465
5 kPa/kg P-C, 10 kPa/kg P-C, (TL)
P-C
Absolute by Vacuum 60 30
G100K 0 100
PG7607,
PG7601,
2465
5 kPa/kg P-C, 10 kPa/kg P-C, (TL)
P-C
Gauge 60 30
BG15K -15 15 FPG8601 FPG8601 Gauge 90 30
1. A reference can be substituted if the recommended reference is unavailable. The substituted reference must be able to cover the MIN and
MAX pressures required for the calibration of the applicable PMM.

Table of Contents

Other manuals for Fluke 6270A

Related product manuals