Calibration with COMPASS for Pressure
See the other guidelines for calibration at the start of this main section. In
addition:
Use the As Found data file to determine the new C0 and C1 calibration
coefficients and calibration date.
Limit traffic in the area during calibration. Make sure that the calibration
bench is free from environmental disturbances such as:
o drastic changes in pressure
o temperature
o humidity
o air flow and vibration from start of calibration to finish.
Calibration must be done in the native operating mode of the PMM. The
native mode is the first letter of the range designation (A = absolute, BG or G
= gauge).
The Product must warm up according to equipment specifications.
Do not turn off or clear the UUT ZOFFSET. Calibration data is collected with
the zero stored in the module.
Gauge (G) and Bi-Directional Gauge (BG) UUTs must be zeroed before data
collection.
Absolute (A) UUTs are not zeroed before data collection.
To use a PG7000 Piston Gauge:
Set up the piston gauge per the instructions in the PG7000 Users Manual.
To use a FPG8601:
Set up the FPG and FPG tools per the instructions in the FPG8601 Users
Manual.