Pressure Controller/Calibrator
Calibration Tables
109
Table 46. PM600- A20M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging
Time [s]
Ascending
Points
1 ATM
1
60 30
2 5000 60 30
3 10 000 60 30
4 15 000 60 30
5 20 000 60 30
Descending
Points
6 15 000 60 30
7 10 000 60 30
8 5000 60 30
9 ATM
1
60 30
1. ATM: atmospheric pressure from laboratory barometer via PG system.
Table 47. BRM600-BA100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 70 60 30
2 80 60 30
3 90 60 30
4 100 60 30
5 110 60 30
Descending
Points
6 100 60 30
7 90 60 30
8 80 60 30
9 70 60 30