6270A
Service Manual
90
2465
The Piston-Cylinder selected must be able to cover the range of 75 kPa to
103 kPa absolute pressure.
PM200 Test Definition Matrices
Table
9. PM200-BG2.5K Pressure Point Matrix
PM200-BG2.5K
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 -2.5
1
90 30
2 -1.5 90 30
3 -0.5 90 30
4 0.5 90 30
5 1.5 90 30
6 2.5 90 30
Descending Points
7 1.5 90 30
8 0.5 90 30
9 -0.5 90 30
10 -1.5 90 30
11 -2.5 90 30
1. UUT is zeroed before setting pressure.