Pressure Controller/Calibrator
Calibration Tables
91
Table 10. PM200-BG35K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 -35
1
60 30
2 -21 60 30
3 -7 60 30
4 7 60 30
5 21 60 30
6 35 60 30
Descending Points
7 21 60 30
8 7 60 30
9 -7 60 30
10 -21 60 30
11 -35 60 30
1. UUT is zeroed before setting pressure.
Table 11. PM200-BG40K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No. Set Pressure
[kPa]
Dwell Time [s] Averaging Time
[s]
Ascending
Points
1 -40
1
60 30
2 -24 60 30
3 -8 60 30
4 8 60 30
5 24 60 30
6 40 60 30
Descending
Points
7 24 60 30
8 8 60 30
9 -8 60 30
10 -24 60 30
11 -40 60 30
1. UUT is zeroed before setting pressure.