6270A
Service Manual
92
Table 12. PM200-BG60K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 -60
1
60 30
2 -36 60 30
3 -12 60 30
4 12 60 30
5 36 60 30
6 60 60 30
Descending
Points
7 36 60 30
8 12 60 30
9 -12 60 30
10 -36 60 30
11 -60 60 30
1. UUT is zeroed before setting pressure.
Table 13. PM200-A100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 2 60 30
2 20 60 30
3 40 60 30
4 60 60 30
5 80 60 30
6 100 60 30
Descending
Points
7 80 60 30
8 60 60 30
9 40 60 30
10 20 60 30
11 2 60 30