Pressure Controller/Calibrator
Calibration Tables
93
Table 14. PM200-BG100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 0
1
60 30
2 20 60 30
3 40 60 30
4 60 60 30
5 80 60 30
6 100 60 30
Descending Points
7 80 60 30
8 60 60 30
9 40 60 30
10 20 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 15. PM200-BG100K Pressure Point Matrix: Negative Pressure
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 0
1
60 30
2 -25 60 30
3 -50 60 30
4 -75 60 30
5 -100
2
60 30
Descending Points
6 -75 60 30
7 -50 60 30
8 -25 60 30
9 0 60 30
1. UUT is zeroed before setting pressure.
2. Pressure point is limited by minimum pressure of reference and ATM pressure.