6270A
Service Manual
94
Table 16. PM200-A200K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 2 60 30
2 40 60 30
3 80 60 30
4 120 60 30
5 160 60 30
6 200 60 30
Descending
Points
7 160 60 30
8 120 60 30
9 80 60 30
10 40 60 30
11 2 60 30
Table 17. PM200-BG200K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 01 60 30
2 40 60 30
3 80 60 30
4 120 60 30
5 160 60 30
6 200 60 30
Descending Points
7 160 60 30
8 120 60 30
9 80 60 30
10 40 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.