Pressure Controller/Calibrator
Control Specifications
7
PM200 Modules
Model Range (SI Units) Range (Imperial Units) Measurement Mode
1-Year
Specification
(%FS)
PM200-BG2.5K -2.5 kPa to 2.5 kPa
-10 inH₂0 to 10 inH₂0
gauge 0.20 %
PM200-BG35K -35 kPa to 35 kPa -5 psi to 5 psi gauge 0.05 %
PM200-BG40K -40 kPa to 40 kPa -6 psi to 6 psi gauge 0.05 %
PM200-A100K 2 kPa to 100 kPa 0.3 psi to 15 psi absolute 0.10 %
PM200-BG100K -100 kPa to 100 kPa -15 psi to 15 psi gauge 0.02 %
PM200-A200K 2 kPa to 200 kPa 0.3 psi to 30 psi absolute 0.10 %
PM200-BG200K -100 kPa to 200 kPa -15 psi to 30 psi gauge 0.02 %
PM200-BG250K -100 kPa to 250 kPa -15 psi to 36 psi gauge 0.02 %
PM200-G400K 0 kPa to 400 kPa 0 psi to 60 psi gauge 0.02 %
PM200-G700K 0 kPa to 700 kPa 0 psi to 100 psi gauge 0.02 %
PM200-G1M 0 MPa to 1 MPa 0 psi to 150 psi gauge 0.02 %
PM200-G1.4M 0 MPa to 1.4 MPa 0 psi to 200 psi gauge 0.02 %
PM200-G2M 0 MPa to 2 MPa 0 psi to 300 psi gauge 0.02 %
PM200-G2.5M 0 MPa to 2.5 MPa 0 psi to 360 psi gauge 0.02 %
PM200-G3.5M 0 MPa to 3.5 MPa 0 psi to 500 psi gauge 0.02 %
PM200-G4M 0 MPa to 4 MPa 0 psi to 580 psi gauge 0.02 %
PM200-G7M 0 MPa to 7 MPa 0 psi to 1000 psi gauge 0.02 %
PM200-G10M 0 MPa to 10 MPa 0 psi to 1500 psi gauge 0.02 %
PM200-G14M 0 MPa to 14 MPa 0 psi to 2000 psi gauge 0.02 %
PM200-G20M 0 MPa to 20 MPa Psi 0 to 3000 psi gauge 0.02 %
Notes
• Gauge mode modules (PM200-GXXX or PM200-BGXXX) with ranges of 100 kPa (15 psi) or greater will support absolute mode
measurement when used with a Barometric Reference Module.
• For temperatures from 15 °C to 18 °C and 28 °C to 35 °C, add 0.003 % FS/°C. Uncertainty for gauge mode modules assumes
routine zeroing. Uncertainty for absolute mode modules includes 1-year zero stability. This specification can be reduced to
0.05 % FS if the PM200 module is zeroed on a continuing basis to remove the 1-year zero stability component. Instrumental
Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference module is
calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
• Instrumental Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference
module is calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
• For temperatures from 15 to 18 C and 28 to 35 C, add 0.003 % FS/C.