EasyManua.ls Logo

Fluke 6270A - PM200 Modules

Fluke 6270A
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pressure Controller/Calibrator
Control Specifications
7
PM200 Modules
Model Range (SI Units) Range (Imperial Units) Measurement Mode
1-Year
Specification
(%FS)
PM200-BG2.5K -2.5 kPa to 2.5 kPa
-10 inH0 to 10 inH0
gauge 0.20 %
PM200-BG35K -35 kPa to 35 kPa -5 psi to 5 psi gauge 0.05 %
PM200-BG40K -40 kPa to 40 kPa -6 psi to 6 psi gauge 0.05 %
PM200-A100K 2 kPa to 100 kPa 0.3 psi to 15 psi absolute 0.10 %
PM200-BG100K -100 kPa to 100 kPa -15 psi to 15 psi gauge 0.02 %
PM200-A200K 2 kPa to 200 kPa 0.3 psi to 30 psi absolute 0.10 %
PM200-BG200K -100 kPa to 200 kPa -15 psi to 30 psi gauge 0.02 %
PM200-BG250K -100 kPa to 250 kPa -15 psi to 36 psi gauge 0.02 %
PM200-G400K 0 kPa to 400 kPa 0 psi to 60 psi gauge 0.02 %
PM200-G700K 0 kPa to 700 kPa 0 psi to 100 psi gauge 0.02 %
PM200-G1M 0 MPa to 1 MPa 0 psi to 150 psi gauge 0.02 %
PM200-G1.4M 0 MPa to 1.4 MPa 0 psi to 200 psi gauge 0.02 %
PM200-G2M 0 MPa to 2 MPa 0 psi to 300 psi gauge 0.02 %
PM200-G2.5M 0 MPa to 2.5 MPa 0 psi to 360 psi gauge 0.02 %
PM200-G3.5M 0 MPa to 3.5 MPa 0 psi to 500 psi gauge 0.02 %
PM200-G4M 0 MPa to 4 MPa 0 psi to 580 psi gauge 0.02 %
PM200-G7M 0 MPa to 7 MPa 0 psi to 1000 psi gauge 0.02 %
PM200-G10M 0 MPa to 10 MPa 0 psi to 1500 psi gauge 0.02 %
PM200-G14M 0 MPa to 14 MPa 0 psi to 2000 psi gauge 0.02 %
PM200-G20M 0 MPa to 20 MPa Psi 0 to 3000 psi gauge 0.02 %
Notes
Gauge mode modules (PM200-GXXX or PM200-BGXXX) with ranges of 100 kPa (15 psi) or greater will support absolute mode
measurement when used with a Barometric Reference Module.
For temperatures from 15 °C to 18 °C and 28 °C to 35 °C, add 0.003 % FS/°C. Uncertainty for gauge mode modules assumes
routine zeroing. Uncertainty for absolute mode modules includes 1-year zero stability. This specification can be reduced to
0.05 % FS if the PM200 module is zeroed on a continuing basis to remove the 1-year zero stability component. Instrumental
Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference module is
calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
Instrumental Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference
module is calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
For temperatures from 15 to 18 C and 28 to 35 C, add 0.003 % FS/C.

Table of Contents