Pressure Controller/Calibrator
Front Panel and Display Assembly
57
Remove the Module Bay Latch Assembly
To remove the module bay latch assembly from the display-panel shell, see
Figure 32:
1. Shut off the supply pressure and vent the system.
2. Remove all pressure lines and fittings from the manifold.
3. Remove the front-panel assembly as explained earlier.
4. Use a T10 Torx to remove the six screws that hold the panel shell to
the front panel.
5. As you separate the display panel shell from the front panel, disconnect the
keyboard-to-display cable .
6. Use a 3 mm hex driver to remove the 6 x 20 shoulder screw , shim , and
wave washer .
7. Remove the module bay latch .