Pressure Controller/Calibrator
Repair and Replacement
25
Repair and Replacement
The subsequent procedures explain how to disassemble parts of the Product. To
reinstall the parts, reverse the steps unless otherwise specified.
Feet and Tilt Stands
To remove or replace the feet and tilt stands, see Figure 10:
1. Shut off the supply pressure and vent the system.
2. Open the door to vent off the pressure between PCM and supply isolation
valve.
3. Make sure the Product is off and disconnect the mains power cord.
4. Remove all pressure lines and fittings from the manifold.
5. Turn over the chassis so the feet and tilt stands are face up.
6. Remove the feet and tilt stands with a 3/16 inch flat-head screw driver.
7. Attach tilt stands to the feet as shown.
8. Install two feet and two tilt stands on the front of the chassis.
9. Install two feet on the rear of chassis.