6270A
Service Manual
104
Table 36. PM600-G200K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 0
1
60 30
2 50 60 30
3 100 60 30
4 150 60 30
5 200 60 30
Descending Points
6 150 60 30
7 100 60 30
8 50 60 30
9 0 60 30
1. UUT is zeroed before setting pressure.
Table 37. PM600-A200K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 10 60 30
2 50 60 30
3 100 60 30
4 150 60 30
5 200 60 30
Descending Points
6 150 60 30
7 100 60 30
8 50 60 30
9 10 60 30