Pressure Controller/Calibrator
Calibration Tables
101
Table 30. PM200-G10M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 0
1
60 30
2 2000 60 30
3 4000 60 30
4 6000 60 30
5 8000 60 30
6 10 000 60 30
Descending
Points
7 8000 60 30
8 6000 60 30
9 4000 60 30
10 2000 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 31. PM200-G14M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 0
1
60 30
2 2800 60 30
3 5600 60 30
4 8400 60 30
5 11 200 60 30
6 14 000 60 30
Descending
Points
7 11 200 60 30
8 8400 60 30
9 5600 60 30
10 2800 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.