Table 28. PM200-G4M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 0
1
60 30
2 800 60 30
3 1600 60 30
4 2400 60 30
5 3200 60 30
6 4000 60 30
Descending
Points
7 3200 60 30
8 2400 60 30
9 1600 60 30
10 800 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 29. PM200-G7M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure
No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time
[s]
Ascending Points
1 0
1
60 30
2 1400 60 30
3 2800 60 30
4 4200 60 30
5 5600 60 30
6 7000 60 30
Descending Points
7 5600 60 30
8 4200 60 30
9 2800 60 30
10 1400 60 30
11 0 60 30