Pressure Controller/Calibrator
Calibration Tables
99
Table 26. PM200-G2.5M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time [s]
Ascending
Points
1 0
1
60 30
2 500 60 30
3 1000 60 30
4 1500 60 30
5 2000 60 30
6 2500 60 30
Descending
Points
7 2000 60 30
8 1500 60 30
9 1000 60 30
10 500 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 27. PM200-G3.5M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time [s]
Ascending
Points
1 0
1
60 30
2 700 60 30
3 1400 60 30
4 2100 60 30
5 2800 60 30
6 3500 60 30
Descending
Points
7 2800 60 30
8 2100 60 30
9 1400 60 30
10 700 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.