6270A
Service Manual
98
Table 24. PM200-G1.4M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time [s]
Ascending
Points
1 0
1
60 30
2 280 60 30
3 560 60 30
4 840 60 30
5 1120 60 30
6 1400 60 30
Descending
Points
7 1120 60 30
8 840 60 30
9 560 60 30
10 280 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 25. PM200-G2M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging
Time [s]
Ascending
Points
1 0
1
60 30
2 400 60 30
3 800 60 30
4 1200 60 30
5 1600 60 30
6 2000 60 30
Descending
Points
7 1600 60 30
8 1200 60 30
9 800 60 30
10 400 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.