Pressure Controller/Calibrator
Calibration Tables
97
Table 22. PM200-G700K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 0
1
60 30
2 140 60 30
3 280 60 30
4 420 60 30
5 560 60 30
6 700 60 30
Descending
Points
7 560 60 30
8 420 60 30
9 280 60 30
10 140 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 23. PM200-G1M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s] Averaging Time [s]
Ascending
Points
1 0
1
60 30
2 200 60 30
3 400 60 30
4 600 60 30
5 800 60 30
6 1000 60 30
Descending
Points
7 800 60 30
8 600 60 30
9 400 60 30
10 200 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.