Pressure Controller/Calibrator
Contamination Prevention System (CPS)
119
10. Inspect to make sure all O-rings are in the correct place prior to re-assembly
(see Figure 50).
The filter retainer has an O-ring on each end.
The filter manifold has 3 O-rings.
The purge tube has one O-ring and a c-clip.
The screen retainer has one O-ring.
The lower housing has one O-ring.
11. Reassemble in reverse order.
hxn032.eps
Figure 50. Pump Inspection