Pressure Controller/Calibrator
Module Calibration
87
Note
Do this in an Excel spreadsheet where C1_Corrected = known_y’s
and reference pressure data = known_x’s.
Write the new calibration date, C0, C1 to the module and clear the zOffset.
1. Write new calibration date to module, enter
CAL_DATE mm/dd/yy
2. Write the new C0 to the module, enter
OFFSET_SET x” where “x” is a decimal or integer in psi.
3. Write the new C1 to the module, enter
GAIN_SET x” where “x” is a decimal (multiplier)
4. Save the values to memory with this command, enter
CAL_STORE
Run As Left data for pressures defined in the calibration matrices at the end of
this manual.