EasyManuals Logo

Fluke 8840A User Manual

Fluke 8840A
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Introduction and Specifications
SPECIFICATIONS
Table
1-1.
Specifications
(eont)
Accuracy,
cont
MEDIUM AND FAST READING RATES
...
In medium rate,add2 counts to thenumberof countsforthe200h
tnrough 200
kn
ranges and 3 counts for the 2000
kn
and 20 Mf)
ranges. In fast reading rate, use3 counts forthe number of count$
for the
200Q
range, and 2 counts for all other ranges.
TEMPERATURE COEFFICIENT
..........
Less than 0.1
x
accuracy specification per OC from O°C to 18*$
and 28OC to 50° C.
MEASUREMENT CONFIGURATION
......
2-wire or 4-wire.
OPEN CIRCUIT VOLTAGE
...............
Less than 6.5V on the 2002) through 200
kfl
ranges. Lessthan 13V
on the 2000
kn
and 20
Ms2
ranges.
INPUT PROTECTION
...................
To 300V rms.
READING
RATES
READING RATES
WITH INTERNAL TRIGGER
..............
(readings per second).
Sensed automatically at power-up.
RATE
S
M
F
AUTOMATIC
SETTLING
TIME
DELAY
POWER LINE FREQUENCY'
1. Difference between first reading and final value for an in-range step change coincident with trigger.
50
HZ
2.08
16.7
100
Time in milliseconds from single trigger to
start
of
A/D
conversion, Autorange off.
2. For slow reading rate.
50
counts for medium rate; 10 counts for
fast
rate.
r
FUNCTION
VDC
VAC
mA DC
mA AC
Ohms
60
HZ
2.5
20
100
400HZ
2.38
19.0
100
NUMBER OF COUNTS
FROM FINAL VALUE1
5
5
30 (Note 2)
5
30
(Note
2)
5
5
5
5
10
30
RANGE
200
mV
2V-10WV
All
2000
mA
2000
mA
200n
2~2
20
ks2
200
kn
2000
kR
20 Mn
READING RATE
F
9
9
551
9
551
17
13
13
21
81
723
S
342
342
551
342
551
394
322
342
141
141
1020
M
61
17
551
17
551
105
17
17
121
101
964
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8840A and is the answer not in the manual?

Fluke 8840A Specifications

General IconGeneral
Frequency Range20 Hz to 100 kHz
Operating Temperature0°C to 50°C
DisplayVacuum Fluorescent
DC Current Range0 to 2 A
AC Current Range0 to 2 A