Safety Instructions
Symbols 1
1-5
Warning
To prevent electrical shock or fire:
• Use only the Fluke power supply, Model BC190 (Power
Adapter).
• Before use, check that the selected/indicated range on the
BC190 matches the local line power voltage and frequency.
• For the BC190/808 and BC190/820 universal Power
Adapters, use line cords that comply with the local safety
regulations.
Note
The BC190/808 and BC190/820 universal Power Adapters are
equipped with a male plug that must be connected to a line cord
appropriate for local use. The adapter is isolated so the line cord
does not need to be equipped with a terminal for connection to
protective ground.
Warning
To prevent electrical shock or fire if a product input is
connected to more than 42 V peak (30 Vrms) or 60 V dc:
• Use only insulated voltage probes, test leads and adapters
supplied with the product, or indicated by Fluke as suitable
for the Fluke 190 Series II ScopeMeter® Test Tool series.
• Before use, inspect voltage probes, test leads, and
accessories for mechanical damage and replace when
damaged.
• Remove all probes, test leads, and accessories that are not
in use.
• Always connect the power adapter first to the ac outlet
before connecting it to the product.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
• Do not connect the ground spring to voltages higher than
42 V peak (30 Vrms) from earth ground.
• When using the ground reference lead with any of the
probes, make sure that the black isolation sleeve is over
the probe tip.
• Do not apply more than the rated voltage, between the
terminals or between each terminal and earth ground.
• Do not apply input voltages above the rating of the
instrument. Use caution when using 1:1 test leads because
the probe tip voltage will be directly transmitted to the
product.
• Do not use exposed metal BNC or banana plug connectors.
Fluke offers cables with plastic, safety designed BNC
connectors suitable for the ScopeMeter
®
Test Tool product.
• Do not insert metal objects into connectors.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com