Chapter 2 MicroLog Plus
148
Note: The password must include at least 4 characters and is case
sensitive.
2. Humidity calibration
To calibrate the humidity sensor you will need a humidity
chamber.
1. Connect the MicroLog to the PC
2. Set up the MicroLog to record every 10 seconds (see
page 131).
3. Click Logger on the menu bar, then click Calibration.
4. Enter the calibration password, then click OK.
5. Select Humidity in the Choose sensor drop list