Chapter 2 MicroLog Plus
150
7. Disconnect MicroLog from the PC and insert it into the
humidity chamber.
8. Set the humidity chamber to the first reference value.
Wait until the humidity level is stabilized and write down
the MicroLog’s reading.
9. Repeat the last step with the second reference value.
10. Connect the MicroLog to the PC.
11. Enter the two MicroLog values into the MicroLog Value
edit boxes.
12. Enter the two reference values into the Reference Value
edit boxes.
13. Click Calibrate.
To restore default calibration, click Default.