Chapter 1 MicroLog
74
humidity chamber.
h. Set the humidity chamber to the first reference value.
Wait until the humidity level is stabilized and write
down MicroLog’s reading.
i. Repeat the last step with the second reference value.
j. Connect MicroLog to the PC.
k. Enter the two MicroLog values into the MicroLog
value edit boxes.
l. Enter the two reference values into the Reference
value edit boxes.
m. Click Calibrate.
To restore default calibration, click Default.