To run this test, follow the on-screen instructions.
Low P leak (machines with ACGO)
The negative Low P Leak measures machine leaks before the
breathing system, between the common gas outlet and the high-
pressure pneumatics and includes the gas mixer and vaporizer. It
measures low pressure pneumatic leaks with a pass or fail limit of 50
ml.
Any cassette may be used during the check. Using a desflurane
cassette only checks for leaks between the vaporizer and common
gas outlet.
To run this test, follow the on-screen instructions.
Agent delivery
The agent delivery test checks the agent delivery systems and the
cassette. This check releases agent to the circuit.
Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
To run this test, follow the on-screen instructions.
Note
Note