3. Set Leak Audio to Off.
The audio alarms for small leaks are silenced.
4.
To turn audio alarms back on, set Leak Audio to On.
5. Select Close.
If the Low MV alarm limits are off or MV/TV Alarms is set to Off,
Leak Audio is automatically set to On and cannot be changed.
--
Setting auto MV limits
MV alarm limits can be calculated automatically for mechanical
ventilation when in VCV or PCV-VG modes and volume
compensation is enabled. Use the Auto MV Limits setting to turn on
automatic calculations of the MV alarm limits.
If the automatic calculation of the low or high minute volume alarm
limit exceeds the allowable limit, the minimum or maximum alarm
limit is used.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Auto MV Limits to On.
The MV alarm limits are automatically calculated until Auto MV
Limits is set to Off or until an MV limit is manually adjusted
during mechanical ventilation. The MV alarm limits are
automatically calculated based on TV and RR settings.
4. Select Close.
Setting to default limits
Use Change to Default Limits to set alarm limits to the values set
by the Super User.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Select Change to Default Limits.
4. Select Close.
Note
Aisys CS²
3-20 2067226-001