Intended use
The Aisys CS
2
is scalable, flexible, and functionally integrated,
featuring the most advanced design, ventilation, respiratory
monitoring, and breathing system. Module bays allow for the physical
integration of legacy Datex-Ohmeda patient monitors and supports
mounting of other GE Healthcare monitors. Optionally, the open
architecture design supports mounting of non-Datex-Ohmeda patient
monitors, record keeping, and connections to the hospital information
system. The INview movable display arm helps keep the
anesthetist’s focus on the patient by offering control of all
hemodynamic, gas delivery, anesthetic agent, and ventilation
parameters.
This anesthesia system is designed for mixing and delivering
inhalation anesthetics, Air, O2, and N2O.
The anesthesia system’s small breathing system volume allows an
anesthetic agent delivery response time of less than 7 seconds.
(Agent response time defined as achieving 90% of the setting
change measured while in non-circle circuit with a 2 l/min fresh gas
flow.)
This anesthesia system uses SmartVent ventilation technology
offering Volume Control Ventilation with tidal volume compensation
and electronic PEEP. The proven SmartVent technology also
features optional Pressure Control Ventilation, Pressure Support
Ventilation with an Apnea Backup (PSVPro) that is used for
spontaneously breathing patients, Synchronized Intermittent
Mandatory Ventilation (SIMV) modes, Pressure Control Ventilation-
Volume Guarantee (PCV-VG), Continuous Positive Airway Pressure
+ Pressure Support Ventilation (CPAP + PSV), and VCV Cardiac
Bypass. In Volume Control Ventilation, a patient can be ventilated
using a minimal tidal volume of 20 ml. In Pressure Control
Ventilation, volumes as low as 5 ml can be measured. These
advanced features allow for the ventilation of a broad patient range.
This system is not suitable for use in an MRI
environment.
Indications for use
The system is intended to provide general inhalation anesthesia and
ventilatory support to a wide range of patients (neonate, pediatric,
and adult). The device is intended for volume or pressure control
ventilation.
General information
This anesthesia system uses the Advanced Breathing System
(ABS). This integrated breathing system is easy to remove and
disassemble and is autoclavable. Its fully integrated design
WARNING
Aisys CS²
1-2 2067226-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE auisys cs2 and is the answer not in the manual?

GE auisys cs2 Specifications

General IconGeneral
BrandGE
Modelauisys cs2
CategoryMedical Equipment
LanguageEnglish

Related product manuals