Suction regulators (optional)
Venturi Suction Regulator
Performance Category Pharyngeal Suction
Supply Air or O2 from system gas supply
Drive Gas Consumption* 28 l/min with pipeline drive gas at 280 kPa
52 l/min with pipeline drive gas at 600 kPa
Maximum Vacuum* 600 mmHg with pipeline drive gas at 280 kPa
550 mmHg with pipeline drive gas at 600 kPa
Maximum Flow* 29 l/min with pipeline drive gas at 280 kPa
32 l/min with pipeline drive gas at 600 kPa
Vacuum Gauge Accuracy +/- 5% of full scale
*Values are approximate.
Continuous Suction Regulator
Performance Category Pharyngeal Suction
Supply External vacuum
Maximum Vacuum* 540 mmHg with external vacuum applied of
540 mmHg and 40 l/min free flow
Maximum Flow* 39 l/min with external vacuum applied of 540
mmHg and 40 l/min free flow
Vacuum Gauge Accuracy +/- 5% of full scale
*Values are approximate.
Aisys CS²
11-22 2067226-001