Auto limits
Use the Auto Limits menu to quickly set alarm ranges for
‘MV’, ‘TV’,
and ‘EtCO2’ during mechanical ventilation.
Setting auto limits
1. Select Auto Limits.
The menu shows the current measured values and the
proposed low and high alarm limits.
2. Check the proposed parameters.
Select Confirm to use the proposed low and high alarm
limits.
Select Cancel to leave the alarm limits unchanged.
Select Case Default Limits to set the alarm limits to the
case default limits.
The proposed low and high alarm limits are shown in highlighted
text. The alarm limits that are not highlighted do not change.
--
Note
3 Operation
2067226-001 3-9