Pneumatic power outlet
The pneumatic outlet is supplied with the same gas supply pressure
as the attached oxygen medical gas pipeline supply. The anesthesia
system is rated for supply pressures between 280 kPa - 600 kPa.
The anesthesia system is designed to function normally with a
maximum flow of 8 l/min from the pneumatic outlet. Higher flows
from this connection are possible and may result in decreased
anesthesia system performance.