Electromagnetic compatibility (EMC)
Changes or modification to this equipment not expressly
approved by the manufacturer could cause EMC issues
with this or other equipment. Contact the manufacturer
for assistance. This device is designed and tested to
comply with applicable regulations regarding EMC as
follows.
Use of portable phones or other radio frequency (RF)
emitting equipment (that exceed electromagnetic
interference levels specified in IEC 60601-1-2) near the
system may cause unexpected or adverse operation.
Monitor operation when RF emitters are in the vicinity.
Use of other electrical equipment adjacent to or stacked
with this system may cause interference. Verify normal
operation of equipment in the system before use on
patients.
Cables
Cables used to obtain system electromagnetic emission and
immunity results are described below.
Use of any cables other than those specified by the
manufacturer may result in increased emissions or
decreased immunity of the anesthesia system.
Display optional output
ports
Cables used for EMC testing
Ethernet Standard 8-pin RJ45 connector with
unshielded Ethernet cable - 7.6 m maximum
length.
Ethernet Standard 8-pin RJ45 connector with
unshielded Ethernet cable - 7.6 m maximum
length.
Universal Serial Bus (USB) Standard Type A connector USB cable with
foil shield - 1.8 m maximum length.
Universal Serial Bus (USB) The visible USB connector can be populated
with a memory stick.
Video Graphics Array (VGA) Standard VGA 15-pin D-subminiature three
row connector with shielded cable - 1.8 m
maximum length.
RS-232 Serial with On/
STBY
Part number M1023823.
WARNING
•
•
WARNING
Aisys CS²
11-46 2067226-001