Changing a cassette during a case
1.
Push the Agent quick key.
2. Set agent delivery to off.
3. Turn the lock on the handle to the vertical position (Aladin2
cassettes only).
4. Remove cassette from the active bay.
• Store the cassette in the cassette storage bay if it will not be
used.
• Fill the cassette following the appropriate filling procedure if
the cassette requires filling.
5. Install a cassette following the Installing a cassette procedure.
A tone sounds.
The agent setting in the quick key area flashes.
If cassette of same agent type is installed during a case, the last
agent setting used shows the agent quick key.
6. Change or confirm the agent setting.
7. To confirm the agent setting:
• Push the ComWheel to confirm the setting.
• Push the Agent quick key to confirm the setting.
If the setting is not confirmed within 30 seconds, the agent
delivery is set to off.
8. To change an agent setting:
• Turn the ComWheel to the desired setting.
• Push the ComWheel or push the quick key to confirm the
setting.
Aisys CS²
13-6 2067226-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE auisys cs2 and is the answer not in the manual?

GE auisys cs2 Specifications

General IconGeneral
BrandGE
Modelauisys cs2
CategoryMedical Equipment
LanguageEnglish

Related product manuals