• Unblock the patient connection.
•
Set the Bag/Vent switch to Vent.
• Set the tidal volume and total flow to minimum.
• Other alarms such as MVexp low can occur.
• Make sure that the Ppeak low. Leak? alarm occurs.
10. Set all alarm limits to approved clinical values.
Aisys CS²
7-16 2067226-001