AB.74p047
5. To clean and disinfect the canister, refer to “Absorbent canister
cleaning” in the “Cleaning and Sterilization” manual.
6. Place a new filter in the bottom of the canister, pour absorbent
into the canister and place a new filter over the absorbent before
closing and locking the cover. Wipe off any absorbent dust.
7. Align the cover slots with the canister locking tabs and press the
cover down into place. Turn the cover locking ring clockwise to
lock the cover in place. Ensure the cover is properly sealed to
prevent leaks and spillage. Alignment of the arrows helps to
indicate correct assembly.
AB.82p001
The filters must be in place to help prevent dust and
particles from entering the breathing circuit.
--
8. When replacing the canister, make sure that it is seated properly
on the support pins or in the EZchange canister module before
latching it into place.
WARNING
8 Setup and connections
2067226-001 8-9