EasyManua.ls Logo

GE auisys cs2

GE auisys cs2
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
11000
12000
13000
14000
15000
18-25C
10-35C
AB.60.056
0
1
3 4
2
+20%, +1.0% v/v
+10%, +0.5% v/v
-10%, -0.5% v/v
-20%, -1.0% v/v
5
6
1. Accuracy is greater of percent of setting (%) or volumetric percent (%
v/v)
2.
Fresh gas flow ml/min
3. All settings
4. Settings equal to or less than 13.4%
5. Settings equal to or less than 9%
6. Settings equal to or less than 3.6%
Figure 11-13 • Desflurane accuracy Aladin2
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
11000
12000
13000
14000
15000
18-25C
10-35C
AB.60.057
1
2
+20%, +0.4% v/v
+10%, +0.2% v/v
-10%, -0.2% v/v
-20%, -0.4% v/v
11 Specifications and theory of operation
2067226-001 11-41

Table of Contents

Related product manuals