Troubleshooting: Wireless LAN
6-28 Dash 3000/4000/5000 2000966-542D
These values should be:
If the association status is other than 01, it is likely 0B (SSID or PASS
KEY incorrect). Verify that the SSID and PASS KEY matches the access
point configuration. Go to “Confirm or configure wireless LAN” on
page 4-14 for steps to check the settings.
If signal strength and transmit rate statuses are less than the above
values, the source of the problem is likely the wireless network (access
point). Go to Analyze WIRELESS LAN STATUS screen below and
contact your local hospital IT department for further troubleshooting.
If association status, signal strength and transmit statuses meet the
requirements, then the wireless patient monitor is not the source of the
problem. Contact your local hospital IT department for further
troubleshooting.
Association Status 01
Received Signal Strength Indicator
(RSSI)
minimum user-defined value
Transmit Rate Status 05 or greater

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output