Equipment overview: Components
2-12 Dash 3000/4000/5000 2000966-542D
Wireless connection
The flexibility of the optional GE CARESCAPE Network is increased by
using the wireless network. The wireless connection allows the user to
roam from one access point to another, maintaining a strong seamless
connection to the CARESCAPE Network. GE offers the 802.11a/b/g
wireless option.
The patient monitor, with its optional built-in wireless card, functionally
performs the same as a patient monitor connected directly to the optional
CARESCAPE Network. It can be viewed at the central station and by
other GE monitors on the network (e.g., Dash 3000/4000/5000, Eagle
4000, and Solar
patient monitors). Patient monitors with a wireless
connection can send and receive patient data via the access points to the
CARESCAPE Network.
NOTE
It is recommended that wireless patient monitors that are moved
from room to room have their patient monitor type configured as
Rover or Rover/Combo monitoring.
To extend the CARESCAPE Network to a hospital’s 802.11a/b/g wireless
network, a proper installation and configuration needs to be performed.
To maintain continuous wireless patient monitoring, refer to the
“Wireless LAN Network Configuration Guide” and contact GE for
consultation in integrating the CARESCAPE Network to a 802.11a/b/g
wireless network.
To identify a patient monitor with the wireless option, look for the
wireless LAN label.
004B
Wireless
LAN label

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable