Placement conditions
Free space required around ImageQuant LAS 4000
Secure space for maintenance work as follows:
: 600 mmFront
: 600 mmRight
: 600 mmLeft
: 200 mmRear
: 1000 mmTop
Table strength
The allowable load must be 981 N/m
2
(100kg/m
2
) or greater.
Installation conditions
Other conditions
1
Decide on an installation location taking into consideration the work flow and
ancillary facilities to be used.
2
Required construction work and electricity/air conditioning work must be complet-
ed in advance.
3
It is not desirable to have a heat source on the IDX right side face where there is
an air intake fan even if the environmental requirements have been met.
4
Do not install the equipment near a window to avoid direct sunlight. Attach a blind
to nearby windows.
5
Do not place objects near the power outlet so that you can disconnect the power
cable from the outlet in case of emergency.
Floor oscillation conditions
: 0.03G (5 to 60 Hz)Oscillation(1) Operating time
: 1GImpact
: 0.4G (5 to 60 Hz)Oscillation(2) Non-operating time
: 2GImpact
130 ImageQuant LAS 4000 User Manual 28-9607-42 AB