IRECTION 5771498-100, REVISION 5 VENUE™ SERVICE MANUAL
5-26 Section 5-10 - System Power Distribution
PRELIMINARY
Section 5-10
System Power Distribution
5-10-1 Introduction
The Venue™ system power distribution consists of the following modules:
• An Isolated AC/DC Power supply to all System modules (18v/400W) called the PSU Unit.
• Optional Battery Packs that supplies Power to the system when AC power line is not present or in
the event of AC power failure (12-16.8V).
• Input filter unit, IPP, before PSU, for filtering AC line noises.
5-10-2 AC Distribution
The PSU module converts the AC power line (100-240V AC) to 18V DC @ 400 W. this module is the
main power supply for the Venue™ ultrasound scanner:
5-10-3 DC Power Distribution
The main DC Power supplied from the PSU (AC/DC) is distributed by the MPB to the following system
modules:
• Back End Module.
• Front End Power Supply (T-FEPS).
• Cockpit (Monitor)
• Battery charging
• Peripherals