8-16 Section 8-3 - Covers - Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-3-8 MPB Door Cover Replacement Procedure
8-3-8-1 Tools
Appropriate Phillips screwdriver
8-3-8-2 Time Required
5 min
8-3-8-3 Preparations
Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
8-3-8-4 MPB Door Cover Removal Procedure
1) Remove all accessories.
2) Remove the following covers: Lower Front eTower cover, RS probe cover, Right Side eTower
Cover and the Left Side eTower Cover:.
3.) Loosen the four captive screws (two on each side), grip the MPB Door Cover at the bottom and pull
toward you to remove it.
FRU Part #
Refer to Table 9-8 on page 9-7.
Accessories - Replacement Procedures
Lower Front eTower Cover Removal Procedure
RS Probe Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Figure 8-8 Removing MPB Door Cover