8-50 Section 8-4 - System Modules- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-9 BEP PCB Replacement Procedure
8-4-9-1 Tools
Appropriate Phillips screwdriver
8-4-9-2 Time Required
15 min
8-4-9-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-9-4 BEP PCB Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower,
3.) Perform BE module removal procedure.
4.) Remove omega bracket: release 5 screws connecting the omega bracket:
5.) Remove the supporting bracket using an appropriate Phillips screwdriver:
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
MPB Front Metal Door Removal Procedure
BE Module Removal Procedure
Figure 8-36 BE Omega Bracket Removal