Chapter 8 - Replacement Procedures 8-43
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-7 Battery Module Replacement Procedure
8-4-7-1 Tools
Appropriate Phillips screwdriver.
8-4-7-2 Time Required
15 min
8-4-7-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-7-4 Battery Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Right Side eTower
Cover, Left Side eTower Cover and MPB Door Cover:.
3.) Release the two captive screws securing each battery pack.
4.) Pull out the top and bottom battery packs.
Refer to Table 9-10 on page 9-10.
• Accessories - Replacement Procedures
• RS Probe Cover Removal Procedure
• Lower Front eTower Cover Removal Procedure
• Right Side eTower Cover Removal Procedure
• Left Side eTower Cover Removal Procedure
• MPB Door Cover Removal Procedure
Figure 8-30 Removing the Battery Pack