8-182 Section 8-8 - Peripherals Replacement/Installation Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-8-5 ECG Replacement Procedure
8-8-5-1 Tools
Appropriate Phillips screwdriver
8-8-5-2 Time Required
15 min
8-8-5-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-5-4 ECG Removal Procedure
1.) Remove the following covers: Lower Front eTower, and Upper Front eTower.
2.) Release the screw that secures the ECG module in place.
Refer to Table 9-20 on page 9-19.
• Lower Front eTower Cover Removal Procedure
• Upper eTower Front Cover Removal Procedure
Figure 8-204 Releasing ECG Securing Screw