IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
7-22 Section 7-4 - Gathering Troubleshooting Data
PRELIMINARY
7-8-7 System Stuck in Stand By Mode--> verify T_CFE Board Revision
Symptom: System Stuck in Stand By Mode
Suggested solution: verify T_CFE board revision
1.) Verify T-CFE board (P/N 5764675_X) revision (X) is 5 or higher (the label is found under the black
metal bracket on the back side of CFE board)
2.) If needed, install R2 302.40.0 SW version (or higher)
Figure 7-44 CFE board revision 5