D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
7-18 Section 7-4 - Gathering Troubleshooting Data
PRELIMINARY
7-8-3 Cockpit ON/OFF Button turns orange -->BEP BIOS Jumper Setup
Symptom: The system does not turn ON (cockpit on/off button turns green for a few seconds and then
turns orange).
Suggested solution: Reset BEP BIOS jumper
1.) Reset BIOS by changing the jumper position as shown in the picture: move from the 2 outer pins to
the 2 inner pins
2.) Turn on the system and then off again
3.) Reset the jumper to its original position
Figure 7-38 Change Jumper Position to Reset BIOS