8.) Press down on the front locking brake to unlock the wheel. Repeat for the back wheel.
9.) Carefully roll the scanner forwards down the unloading ramp to remove it completely from the
shipping carton.
10.)Make sure the system is standing firmly on the floor before applying the locking brake.
11.)Before commencing the installation procedure, follow the instructions for Physical Inspection on
page 3 - 17.
12.)Remove the packing slip (shipping consignment note) detailing the contents of the shipping carton
and keep close to hand ready to mark the check list - refer to Physical Inspection on page 3 - 18.
Figure 3-14 Removing Silica Gel Bags